FAQs

Q?

Why is the display blinking?

A.

-Sensor output out of specified range

Solution: recalibration

On VOD: the 4mA depends on the pipe diameter but must be calibrated only on pure water, the 20mA depends on the maximum range specified by the process. The sensor maximum range equals to cloaked lens and the FU is 10000 (ESC/OUTPCONF/USERPNTS/POINT2/EDIT/10000/ENTER and save)

on VOM: the 4mA equals to pure water and 20mA equals to 2% Fat milk. This range can be changed thanks to the calibration menu

ESC/CALIB/OK?/4mA(water)/20mA(full product)/DONE

on VOF: The 4mA is usually for pure water and the 20mA for cloaked lens.

Cloacked lens on VOF equals to 1500NTU

Q?

How do I zero the VOM, VOD and VOF on water?

A.

Download (PDF, 159KB)

Q?

What is the VO program structure?

A.

Vo turbidity meter program structure

Q?

How do I calibrate a VO turbidity meter on 2 points?

A.

2 POINTS CALIBRATION PROCEDURE:

Download (PDF, 165KB)